A modified cumulative sum technique has been applied to radioimmunoassay quality control data. The method is approximately 50% more efficient in detecting systematic changes in the mean and variance of quality control values for plasma samples than more widely used conventional methods. The salient features of the technique have been restricted to changes in the mean quality control value of a plasma pool, but potential applications to changes in variance and as a diagnostic aid to problems in radioimmunoassay have been evaluated. The method is independent of computing facilities and statistical expertise since all computations have been presented in the form of a nomogram and thus can be used by technicians at the bench.
Journal of Endocrinology is committed to supporting researchers in demonstrating the impact of their articles published in the journal.
The two types of article metrics we measure are (i) more traditional full-text views and pdf downloads, and (ii) Altmetric data, which shows the wider impact of articles in a range of non-traditional sources, such as social media.
More information is on the Reasons to publish page.
Sept 2018 onwards | Past Year | Past 30 Days | |
---|---|---|---|
Full Text Views | 3 | 0 | 0 |
PDF Downloads | 3 | 1 | 0 |